| Configuration | |
| Model | SGA-506 (2-in-1 configuration; 4-in-1 and 6-in-1 also available) |
| Type | Industrial fixed, explosion-proof (flameproof) |
| Detectable Gases | 1–2 gases from 1,000+ combustible / toxic / VOC types; extendable to dust, noise, temperature, humidity |
| Sensing Technologies | Electrochemical, infrared, catalytic combustion, PID, semiconductor, optical waveguide, MOS |
| Detection Ranges | 0–1 / 2 / 5 / 10 / 50 / 100 / 500 / 1000 / 5000 / 10000 / 50000 (gas and technology dependent) |
| Resolution | 0.001 / 0.01 / 0.1 / 1 / 2 / 5 (gas and technology dependent) |
| Units | PPM, PPB, %LEL, %VOL, mg/m³, μg/m³, g/m³, g/l, mol/mol, μmol/mol |
| Performance | |
| Accuracy | ≤±3% FS (gas dependent) |
| Repeatability | ≤±1% FS (gas dependent) |
| Linearity Error | ≤±1% FS (gas dependent) |
| Response Time | ≤3–20 s (gas dependent) |
| Sampling Method | Natural diffusion (default); pipeline / flow-through / pump-sampling optional |
| Electrical | |
| Operating Voltage | DC 24V (12–30V) |
| Power Consumption | Electrochemical ≤1.0W; IR / catalytic / laser TDLAS / PID / semiconductor / optical waveguide / MOS ≤1.5W |
| Output Signals | RS485, RS232, 4-20mA, 0-5V, Ethernet; WIFI, LoRa, ZigBee, NB-IoT, GPRS, Bluetooth (wireless) |
| Relay Output (optional) | Three-level alarm / two sets of 1A/24VDC normally-open dry contacts (normally-closed optional) |
| Environmental | |
| Operating Temperature | −20℃ ~ +50℃ (up to 1200℃ with high-temperature sampling kit) |
| Operating Humidity | 10–95% RH (up to 99% RH with dehumidification kit) |
| Operating Pressure | 86–106 kPa |
| Mechanical | |
| Dimensions (2-gas) | 154 × 278 × 142 mm (without local alarm) / 206 × 278 × 142 mm (with local alarm) |
| Weight (2-gas) | 2.3 kg (without local alarm) / 2.5 kg (with local alarm) |
| Housing Material | Die-cast aluminum alloy |
| Display | 2.4-inch LCD color screen |
| Alarm (optional) | Sound (≥85 dB) + strobe light + on-screen text; high/low alarm, fault, calibration & sensor-life reminders |
| Electrical Interface | M20×1.5 (G3/4 thread optional) |
| Protection Rating | IP66 |
| Certifications & Standards | |
| Explosion Protection | Ex db IIC T6 Gb |
| Certifications | CPA type approval, national explosion-proof certificate, SIL, CE, ROHS, ATEX, third-party metrology certificate |
| Design Standards | GB50493-2019; GB12358-2024; JJF 1368-2012; GB/T3836.1-2021; GB/T3836.2-2021; GB/T20936.1-2022; GBZ 2.1-2019; GB/T4208-2008 |
| Functions | |
| Smart Calibration | Zero auto-calibration, multi-point span calibration, baseline auto-correction |
| Data Storage (optional) | 100,000+ local records, 1–999s interval, USB export; unlimited IoT cloud storage |
| Accessories | Packaging box, manual, certificate, shipping list |
| Options | IR remote control, temperature/humidity module (−40~+80℃ / 0–100%RH), Bluetooth printer |