1. Product Overview
The GASSHIELD UGT-E4-HCL is a high-performance fixed-potential electrolytic (electrochemical) hydrogen chloride gas sensor designed for accurate and reliable detection of hydrogen chloride (HCl) in industrial and environmental applications. Utilizing a three-electrode electrochemical cell architecture, the sensor delivers a detection range of 0–100 ppm with 2 ppm resolution and a sensitivity of 200 ± 100 nA/ppm. With T90 < 70 s response time, excellent repeatability (2% of output signal), and long-term stability of <2% signal per month, the UGT-E4-HCL provides precise HCl monitoring for high-concentration, safety-critical environments.
The sensor operates at a low +200 mV bias voltage and supports a wide load resistance range of 5–30 Ω. Housed in a compact Φ20.2 mm form factor, the UGT-E4-HCL is ideal for integration into portable gas detectors, fixed gas alarm systems, chemical process monitoring equipment and industrial safety instruments where hydrogen chloride leak detection and occupational safety are paramount.
2. Technical Principle
The UGT-E4-HCL employs a fixed-potential electrolytic (electrochemical three-electrode) detection principle. The sensor cell consists of a working electrode, a counter electrode, and a reference electrode immersed in an electrolyte solution separated from the ambient atmosphere by a gas-permeable membrane.
When hydrogen chloride gas diffuses through the membrane and reaches the working electrode, it undergoes a reaction, releasing electrons and generating a current. Simultaneously, a corresponding reaction occurs at the counter electrode. The magnitude of the current flowing between the working and counter electrodes is directly proportional to the hydrogen chloride concentration in the surrounding atmosphere. This current is converted to a voltage signal via an external load resistor (5–30 Ω) or transimpedance amplifier circuit, enabling precise quantification of HCl concentration from 0 to 100 ppm.
The reference electrode maintains a stable potential at the working electrode, ensuring consistent sensitivity and minimizing drift over the sensor's 2-year service life. The +200 mV bias voltage optimizes the electrochemical reaction for HCl while suppressing unwanted background currents.
3. Key Features & Competitive Advantages
A. High Accuracy & Long Service Life
B. Fast Response, Quick Return to Zero
-
T90 < 70 s for rapid detection of hydrogen chloride leaks and hazardous exposure.
-
Quick recovery to baseline after gas exposure, enabling continuous monitoring cycles.
C. Low Power Consumption, High Sensitivity
D. Wide Linear Range, Strong Anti-Interference
-
Linear output characteristic simplifies calibration and signal processing.
-
Minimal response to common gases: CO (−1 ppm), NO₂ (−1 ppm).
E. Excellent Repeatability & Stability
-
Repeatability within 2% of output signal for consistent measurement precision.
-
Long-term stability <2% signal per month, reducing frequent recalibration requirements.
F. Wide Range of Application Fields
-
Suitable for industrial safety, chemical processing, semiconductor manufacturing, and environmental monitoring.
G. Compact Size, Cost-Effective
4. Technical Specifications
Note: Cross-sensitivity may vary by ±30%. Calibration with target gas is recommended. This sensor is NOT suitable for environments containing nitric oxide or hydrogen sulfide.
5. Applications
The GASSHIELD UGT-E4-HCL is engineered for versatile deployment across hydrogen chloride detection and safety monitoring scenarios:
-
Hydrogen Chloride Gas Detectors: Core sensing element for portable and fixed HCl detection instruments in high-concentration environments.
-
Chemical Process Monitoring: HCl concentration measurement in chlorination, PVC production and chemical synthesis plants.
-
Industrial HCl Detection: Leak detection in chemical plants, steel pickling lines and industrial processes.
-
Industrial Safety: Fixed and portable gas detection systems for worker protection in HCl handling areas.
-
Gas Leak Detection: Early warning systems for hydrogen chloride storage tanks, pipelines and distribution equipment.
-
Steel and Chemical Industry: Safety monitoring in pickling lines and acid-based chemical processes.
-
Biomedical: Hydrogen chloride detection in biomedical research and laboratory applications.
-
Smart Municipal: Municipal industrial safety monitoring for smart city projects.
6. OEM Integration & Design Guidelines
To maximize design flexibility and ensure optimal sensor performance, the UGT-E4-HCL is supported by comprehensive integration resources and handling guidelines for equipment manufacturers.
-
Basic Circuit Reference: Standard transimpedance amplifier circuit with R_gain and R_load provided for immediate PCB design integration. Compatible with 5–30 Ω load resistor configurations. Note: a +200 mV bias voltage must be applied between the reference and working electrodes — circuit designs assuming zero bias are not compatible.
-
Temperature Compensation: Baseline output remains within approximately ±1 ppm equivalent from −20°C to +30°C, rising to approximately +6 ppm equivalent at +50°C. Sensitivity varies from approximately 90% to 125% of the 20°C value across −20°C to +50°C. Firmware temperature compensation is recommended for applications operating outside the 0–30°C range.
-
Response & Recovery Characterization: Verified response and recovery test data (50 ppm HCl) enables alarm threshold optimization and adaptive filtering in host firmware.
-
Handling Precautions:
-
Sensor pins must be connected via a PCB socket; soldering or pin bending will damage the sensor.
-
During storage, the working electrode and reference electrode should be kept short-circuited.
-
Avoid contact with organic solvents, alcohol, paints, oils, silicone adhesives, and high-concentration gases.
-
Calibrate and test with standard gas using air as the background gas; oxygen participation is required for the electrochemical reaction.
-
Avoid prolonged use in corrosive gas environments.
-
Non-standard testing methods (concentrated ammonia water, cigarette smoke, lighter gas, exhaled breath, alcohol vapors) will expose the sensor to damaging concentrations — always use target gas with air as the background.
-
Calibration Recommendation: Calibration with the target gas is recommended; cross-sensitivity may vary by ±30%. Calibration should be performed in clean atmosphere with stable, gentle gas flow to simulate a diffusion state.