GASSHIELD UGT-E4-H2S Hydrogen Sulfide Gas Sensor (0-100ppm)

No.UGT-E4-H2S-2
Electrochemical three-electrode hydrogen sulfide gas sensor. 0-100ppm range, 650±200nA/ppm, 1ppm resolution, T90<30s. Φ20.2mm, zero bias, 2-year life.
  • GASSHIELD UGT-E4-H2S Hydrogen Sulfide Gas Sensor (0-100ppm)
  • GASSHIELD UGT-E4-H2S Hydrogen Sulfide Gas Sensor (0-100ppm)
  • GASSHIELD UGT-E4-H2S Hydrogen Sulfide Gas Sensor (0-100ppm)
  • GASSHIELD UGT-E4-H2S Hydrogen Sulfide Gas Sensor (0-100ppm)
  • GASSHIELD UGT-E4-H2S Hydrogen Sulfide Gas Sensor (0-100ppm)
  • GASSHIELD UGT-E4-H2S Hydrogen Sulfide Gas Sensor (0-100ppm)

Description

1. Product Overview

The GASSHIELD UGT-E4-H2S is a high-performance fixed-potential electrolytic (electrochemical) hydrogen sulfide gas sensor designed for accurate and reliable detection of hydrogen sulfide (H₂S) in industrial and environmental applications. Utilizing a three-electrode electrochemical cell architecture, the sensor delivers a detection range of 0–100 ppm with 1 ppm resolution and a sensitivity of 650 ± 200 nA/ppm. With T90 < 30 s response time, excellent repeatability (1% of output signal), and long-term stability of <2% signal per month, the UGT-E4-H2S provides precise H₂S monitoring for high-concentration, safety-critical environments.
The sensor operates without bias voltage and supports a wide load resistance range of 5–30 Ω, simplifying circuit design integration. Housed in a compact Φ20.2 mm form factor, the UGT-E4-H2S is ideal for integration into portable gas detectors, fixed gas alarm systems, and industrial process control instruments where hydrogen sulfide leak detection and occupational safety are paramount.

2. Technical Principle

The UGT-E4-H2S employs a fixed-potential electrolytic (electrochemical three-electrode) detection principle. The sensor cell consists of a working electrode, a counter electrode, and a reference electrode immersed in an electrolyte solution separated from the ambient atmosphere by a gas-permeable membrane.
When hydrogen sulfide gas diffuses through the membrane and reaches the working electrode, it undergoes an oxidation reaction, releasing electrons and generating a current. Simultaneously, a corresponding reduction reaction occurs at the counter electrode. The magnitude of the current flowing between the working and counter electrodes is directly proportional to the hydrogen sulfide concentration in the surrounding atmosphere. This current is converted to a voltage signal via an external load resistor (5–30 Ω) or transimpedance amplifier circuit, enabling precise quantification of H₂S concentration from 0 to 100 ppm.
The reference electrode maintains a stable potential at the working electrode, ensuring consistent sensitivity and minimizing drift over the sensor's 2-year service life. Zero bias voltage operation reduces power consumption and simplifies interfacing with microcontroller-based acquisition systems.

3. Key Features & Competitive Advantages

A. High Accuracy & Long Service Life

  • Detection accuracy with excellent linear response (R² = 0.999) across 0–100 ppm range.
  • 2-year service life in typical operating conditions with 1-year warranty.

B. Fast Response, Quick Return to Zero

  • T90 < 30 s for rapid detection of hydrogen sulfide leaks and hazardous exposure.
  • Quick recovery to baseline after gas exposure, enabling continuous monitoring cycles.

C. Low Power Consumption, High Sensitivity

  • Zero bias voltage operation minimizes power draw — ideal for battery-powered portable instruments.
  • High sensitivity of 650 ± 200 nA/ppm ensures reliable detection across the full measurement range.

D. Wide Linear Range, Strong Anti-Interference

  • Linear output characteristic (R² = 0.999) simplifies calibration and signal processing.
  • Strong anti-interference capability with minimal response to common interfering gases (see cross-interference table).

E. Excellent Repeatability & Stability

  • Repeatability within 1% of output signal for consistent measurement precision.
  • Long-term stability <2% signal per month, reducing frequent recalibration requirements.

F. Wide Range of Application Fields

  • Suitable for industrial safety, petrochemical, oil & gas, wastewater treatment, biomedical, and smart municipal monitoring.

G. Compact Size, Cost-Effective

  • Φ20.2 mm compact form factor enables integration into space-constrained instruments.
  • Cost-effective solution for OEM manufacturers developing hydrogen sulfide detection products.

H. Wide Operating Environment

  • Operating temperature: -30°C to +50°C.
  • Operating humidity: 15–90%RH (non-condensing).
  • Operating pressure: 90–110 kPa.

4. Technical Specifications

Parameter Specification
Detection Performance
Target Gas Hydrogen Sulfide (H₂S)
Detection Principle Electrochemical three-electrode
Detection Range 0 ~ 100 ppm
Max. Overload 500 ppm
Sensitivity 650 ± 200 nA/ppm
Resolution 1 ppm
Zero Drift ±2 ppm
Linearity Linear, R² = 0.999
Repeatability 1% of output signal
Long-Term Stability <2% signal / month
Response Time (T90) <30 s
Service Life 2 years
Warranty 1 year
Operating Environment
Operating Temperature -30 ~ +50 °C
Operating Humidity 15 ~ 90%RH (non-condensing)
Operating Pressure 90 ~ 110 kPa
Electrical Characteristics
Bias Voltage 0
Load Resistance 5 ~ 30 Ω
Physical
Dimensions Φ20.2 mm
Cross Interference
Carbon Monoxide (CO) @ 300 ppm 1 ppm H₂S equivalent
Sulfur Dioxide (SO₂) @ 20 ppm 1 ppm H₂S equivalent
Ammonia (NH₃) @ 50 ppm 0 ppm H₂S equivalent
Nitrogen Dioxide (NO₂) @ 20 ppm -1 ppm H₂S equivalent
Hydrogen (H₂) @ 1000 ppm 0 ppm H₂S equivalent
Nitric Oxide (NO) @ 50 ppm 2 ppm H₂S equivalent
Chlorine (Cl₂) @ 20 ppm -1.4 ppm H₂S equivalent

5. Applications

The GASSHIELD UGT-E4-H2S is engineered for versatile deployment across hydrogen sulfide detection and safety monitoring scenarios:
  • Instrumentation: Core sensing element for hydrogen sulfide gas analyzers and detection instruments.
  • Oil & Gas: H₂S monitoring in drilling operations, refineries, sour gas processing, and petrochemical plants.
  • Industrial Safety: Fixed and portable gas detection systems for worker protection in confined spaces and high-concentration H₂S areas.
  • Gas Leak Detection: Early warning systems for hydrogen sulfide release in industrial processes.
  • Wastewater Treatment: Sewer and drainage H₂S monitoring for odor and toxic gas control.
  • Smart Municipal: Municipal sewer network hydrogen sulfide detection for smart city projects.
  • Biomedical: Hydrogen sulfide detection in biomedical research and medical instrumentation.
  • Environmental Monitoring: Indoor air quality and atmospheric H₂S monitoring.

6. OEM Integration & Design Guidelines

To maximize design flexibility and ensure optimal sensor performance, the UGT-E4-H2S is supported by comprehensive integration resources and handling guidelines for equipment manufacturers.
  • Basic Circuit Reference: Standard transimpedance amplifier circuit with R_gain and R_load provided for immediate PCB design integration. Compatible with 5–30 Ω load resistor configurations and zero-bias operation.
  • Temperature Compensation: Baseline temperature dependence curve (-20°C to +50°C) and sensitivity temperature dependence curve (-20°C to +50°C, ~82%–107% of 20°C value) provided for software compensation algorithms, ensuring accuracy across operating temperature extremes.
  • Response & Recovery Characterization: Full response and recovery curve data (tested at 25 ppm H₂S) enables alarm threshold optimization and adaptive filtering in host firmware.
  • Handling Precautions:
    • Sensor pins must be connected via PCB socket; soldering or pin bending will damage the sensor.
    • During storage, working electrode and reference electrode should be kept short-circuited.
    • Avoid contact with organic solvents, alcohol, paints, oils, silicone adhesives, and high-concentration gases.
    • Calibrate and test with standard gas using air as background gas; oxygen participation is required for electrochemical reaction.
    • Avoid prolonged exposure to corrosive gases and non-standard testing methods (direct ammonia water exposure, cigarette smoke, lighter gas, exhaled breath, alcohol vapors).
  • Calibration Recommendation: Target gas calibration is recommended; cross-sensitivity may vary by ±30%. Calibration should be performed in clean atmosphere with stable, gentle gas flow to simulate diffusion state.