1. Product Overview
The GASSHIELD UGT-E4-CL2 is a high-performance fixed-potential electrolytic (electrochemical) chlorine gas sensor designed for accurate and reliable detection of chlorine (Cl₂) in industrial and environmental applications. Utilizing a three-electrode electrochemical cell architecture, the sensor delivers a detection range of 0–100 ppm with 1 ppm resolution and a sensitivity of −450 ± 150 nA/ppm. With T90 < 60 s response time, excellent repeatability (2% of output signal), and long-term stability of <2% signal per month, the UGT-E4-CL2 provides precise chlorine monitoring for high-concentration, safety-critical environments.
The sensor operates without bias voltage and supports a wide load resistance range of 5–30 Ω, simplifying circuit design integration. Housed in a compact Φ20.2 mm form factor, the UGT-E4-CL2 is ideal for integration into portable gas detectors, fixed gas alarm systems, air purification equipment and industrial process control instruments where chlorine leak detection and occupational safety are paramount.
2. Technical Principle
The UGT-E4-CL2 employs a fixed-potential electrolytic (electrochemical three-electrode) detection principle. The sensor cell consists of a working electrode, a counter electrode, and a reference electrode immersed in an electrolyte solution separated from the ambient atmosphere by a gas-permeable membrane.
When chlorine gas diffuses through the membrane and reaches the working electrode, it undergoes a reduction reaction, releasing electrons and generating a current. Simultaneously, a corresponding oxidation reaction occurs at the counter electrode. The magnitude of the current flowing between the working and counter electrodes is directly proportional to the chlorine concentration in the surrounding atmosphere. This current is converted to a voltage signal via an external load resistor (5–30 Ω) or transimpedance amplifier circuit, enabling precise quantification of Cl₂ concentration from 0 to 100 ppm.
The reference electrode maintains a stable potential at the working electrode, ensuring consistent sensitivity and minimizing drift over the sensor's 2-year service life. Zero bias voltage operation reduces power consumption and simplifies interfacing with microcontroller-based acquisition systems.
3. Key Features & Competitive Advantages
A. High Accuracy & Long Service Life
B. Fast Response, Quick Return to Zero
-
T90 < 60 s for rapid detection of chlorine leaks and hazardous exposure.
-
Quick recovery to baseline after gas exposure, enabling continuous monitoring cycles.
C. Low Power Consumption, High Sensitivity
D. Wide Linear Range, Strong Anti-Interference
-
Linear output characteristic simplifies calibration and signal processing.
-
Strong anti-interference capability with minimal response to common gases: CO (0 ppm), H₂ (0 ppm), H₂S (−1 ppm).
E. Excellent Repeatability & Stability
-
Repeatability within 2% of output signal for consistent measurement precision.
-
Long-term stability <2% signal per month, reducing frequent recalibration requirements.
F. Wide Range of Application Fields
-
Suitable for industrial safety, chemical processing, water treatment, air purification, and environmental monitoring.
G. Compact Size, Cost-Effective
H. Wide Operating Environment
-
Operating temperature: −30°C to +50°C.
-
Operating humidity: 15–90%RH (non-condensing).
-
Operating pressure: 90–110 kPa.
4. Technical Specifications
Note: Cross-sensitivity may vary by ±30%. Calibration with target gas is recommended.
6. Applications
The GASSHIELD UGT-E4-CL2 is engineered for versatile deployment across chlorine detection and safety monitoring scenarios:
-
Instrumentation: Core sensing element for chlorine gas analyzers and detection instruments.
-
Industrial Chlorine Detection: Chlorine leak detection in chlor-alkali plants, chemical synthesis and process industries.
-
Industrial Safety: Fixed and portable gas detection systems for worker protection in chlorine handling areas.
-
Gas Leak Detection: Early warning systems for chlorine storage tanks, pipelines, and distribution equipment.
-
Air Purification: Monitoring chlorine levels in HVAC systems, clean rooms, and indoor air quality management.
-
Water Treatment: Chlorine gas detection at water disinfection plants, swimming pools, and wastewater treatment facilities.
-
Biomedical: Chlorine detection in biomedical research and medical instrumentation.
-
Smart Municipal: Municipal water and sanitation chlorine monitoring for smart city projects.
7. OEM Integration & Design Guidelines
To maximize design flexibility and ensure optimal sensor performance, the UGT-E4-CL2 is supported by comprehensive integration resources and handling guidelines for equipment manufacturers.

-
Temperature Compensation: Baseline temperature dependence curve (−20°C to +50°C) and sensitivity temperature dependence curve (−20°C to +50°C, ~85%–100% of 20°C value) provided for software compensation algorithms — see Section 5.2 / 5.3.
-
Response & Recovery Characterization: Full response and recovery curve data enables alarm threshold optimization and adaptive filtering in host firmware — see Section 5.1.
-
Handling Precautions:
-
Sensor pins must be connected via a PCB socket; soldering or pin bending will damage the sensor.
-
During storage, the working electrode and reference electrode should be kept short-circuited.
-
Avoid contact with organic solvents, alcohol, paints, oils, silicone adhesives, and high-concentration gases.
-
Calibrate and test with standard gas using air as the background gas; oxygen participation is required for the electrochemical reaction.
-
Avoid prolonged use in corrosive gas environments.
-
Non-standard testing methods (concentrated ammonia water, cigarette smoke, lighter gas, exhaled breath, alcohol vapors) will expose the sensor to damaging concentrations — always use target gas with air as the background.
-
Calibration Recommendation: Calibration with the target gas is recommended; cross-sensitivity may vary by ±30%. Calibration should be performed in clean atmosphere with stable, gentle gas flow to simulate a diffusion state.